ICP - IEPE- Senosor Supplys
NeIEPE-8: ICP®-Supply
NeIEPE-8 is an 8-channel IEPE / ICP® supply with two isolated outputs per channel.
- Two isolated outputs per channel
- The IEPE supply provides the required supply energy for the electronics integrated in the sensor, decouples the measuring signal from the output signal of the sensor and provides the measuring signal at the two outputs for further processing.
- Due to the selected circuit, the signal ground of the outputs are connected to the signal ground of the input via 10 KOhm each.
- The maximum voltage difference between input signal ground and output signal ground may be max. 10 Volt.
- Power is supplied to the module via a plug-in power supply.
- The input has reverse polarity protection but no overvoltage protection./li>
- The built-in DC / DC converter has a potential separation of 1 KV.
|
8 | |
Connection | BNC | |
Dimensions | LxWxH: 95 x 112 x 22 mm | |
Weight |
about. 1000 g | |
Storage temperature | -20 .. +70 °C | |
Operation temperature | 0..+50 °C | |
Supply voltage | U VCC = 10 .. 12 VDC | |
Current consumption |
IVCC = typ. 110 mA DC @ with 8 sensors |
|
Protection |
Available |
|
Overvoltage protection | Not available | |
Potential separation | DC / DC converter with 1 kVRMS | |
IEPE/ICP®-supply voltage | UICP = 24 VDC | |
Constant current | Iconst = 4.0 mA +- 0.5 mA | |
Lower limit frequency | FUG = 1 Hz (einpoliger Hochpass, -3dB) | |
Upper cutoff frequency | FOG = 50 KHz (-3dB) | |
Amplification factor |
|
|
Output voltage | +- 10V | |
Output voltage | +- 10V |
The in-house development of this module has the advantage that we can adapt it to the requirements of our customers by adapting the software or hardware. We would be happy to offer you a special solution for your problem.
You can find more information in the user manual.
The prices of the modules can be found in our price list.
NeIEPE-4-DIN 4 channel IEPE-Supply
The NeIEPE-4-DIN module is a 4-channel module for supplying IEPE sensors. It provides the necessary supply energy for the electronics integrated in the sensor and decouples the measurement signal from the sensor output signal. The NeIEPE-4-DIN is built into a housing for DIN rail mounting. The connection is made via screwless terminals.
- Number of channels = 4
- Assembly = DIN Rail TS35
- Connection = screw terminals, max. 1.5 sqm
- Constant current = 4.5 mA + - 0.5 mA
- Power supply = 2 variants
- Direct power supply, without DC / DC converter, without electrical isolation
- Power supply with DC / DC converter
You can find more information in the data sheet. and the user manual.